Simulation of|carrier profiles For measurements at thick samples like ingots, it is very important to simulate the carrier depth profile, that developes in the sample. The measured lifetime is strongly effected by this profile, so that this effect has to be taken into account.
The measurement at thick samples as ingots arises new questions and problems. One of these questions is how the carrier profiles that develop in a sample effect the lifetime measurements. To solve this problem, we developed a simulation tool for the simulation of carrier profiles. This tool consists of a partial differential equation system of the transport equations for electrons and holes and the Poisson equation.
![\frac{\partial}{\partial t}n(x,t)=\frac{\partial}{\partial x}\left[-\mu_nn(x,t)\frac{\partial}{\partial x}\Psi (x,t)+D_n\frac{\partial}{\partial x}n(x,t)\right]+G^o(x,t)-U(x,t)](typo3temp/math/2fe75be883aa62ed09e4a91e654d6d89.png)
![\frac{\partial}{\partial t}p(x,t)=\frac{\partial}{\partial x}\left[\mu_p p(x,t)\frac{\partial}{\partial x}\Psi (x,t)+D_p\frac{\partial}{\partial x}p(x,t)\right]+G^o(x,t)-U(x,t)](typo3temp/math/30ee0243c69929163a7c2be0a012bd79.png)
![\frac{\partial^2}{\partial x^2}\Psi(x,t)=-\frac{q}{\epsilon_0\epsilon_y}\left[-n(x,t)+p(x,t)\pmN_{dot}\right]](typo3temp/math/28a39a028d316d7c2a656fffcdba6e13.png)
Transport equations for electrons and holes + Poisson equation
The simulations for measurements with a long (typical MDP condition) or a very short light pulse (typical µ-PCD conditions) at thick unpassivated samples are shown in figure 1 and 2. It becomes clear, that the carrier profile of a long light pulse expands through a large volume of the sample, where as the carrier profile of the only 200 ns long light pulse is very surface near. This has a direct effect on the measured lifetime, since the surface recombination has a much stronger effect on the µ-PCD measurements.
Figure 3 shows the quantitative effect of the surface recombination on the effective lifetime for both measurement conditions. The MDP measurements are less prone to the surface effect, so that MDP measurements are more suited for the investigation of bulk properties. Accordingly µ-PCD is an ideal method to investigate the surface properties of a sample.